User Manual for PC-80D Easy ECG Monitor
17
7 Measurement option setting
On the default menu,select the “Option”,then press the “”ok key to enter into the optional parameters setting for measurement, as shown in figure
7-1.
On-screen display:
User ID: Displaying the established ID No. or “New ID”. At most 100 ID can be created. When creating a
new ID, at most 30 characters can be entered; the default ID is “00000”, the number of “00000” ~ “00004”
is reserved (already created at factory) and can not be deleted.
Mode: it is valid only for external lead wires. “Limb lead” and “Chest lead” are optional. The “Limb lead”
is external electrodes measurement with 4 lead wires; the “Chest lead” is the external electrodes
measurement with 3 lead wires, it is intended for professional use (ie. by doctor). The “limb lead” or quick
measurement by built-in electrodes is recommended for normal use. The default setting is “limb lead”.
Chest lead: when the measuring mode is set to “chest lead”, V lead is defaulted setting. If you want to obtain specific Chest Lead (V1-V6), select the
lead name (V1-V6) from the pull-down menu.
Period: “30S”, “1min”, “5min”, “10min”, “30min”, “continuous” are optional. The default measuring time is “continuous”.
HR alarm
“On “and “Off “are optional. The default setting is “On”.
HR Hi/Lo: 35~240bpm, the default value is 120bpm; the low alarm limit setting range: 30~235bpm, the default value is 60 bpm.
operation
1. Press the navigation key “/, or ”: Move the cursor upwards/backwards.
2.Press“” ok key:Conform to option and enter into the corresponding parameter setting.
3. Press the navigation key “ or ”: Adjust parameter values.
4. Press“” ok key:Conform the selection, if you do not want to adjust parameter values, or the setting is accomplished,please enter the“
Back”
button.