Conditions That May Affect Use
Environmental
Fluctuating or flickering light may affect measurements.
Use of the 950 outside stated temperature and humidity
specifications may affect measurements. Refer to page 1,
Contraindications, Warnings, and Cautions.
Motion
Excessive or rapid movement may affect measurements.
Sensor Position
Sensor position is critical for accurate SpO
2
measurements.
All emitted sensor light must pass through the finger before
reaching the detector. Do not allow positioning such that
some or all of the emitted light can by-pass the finger as
this will result in inaccurate SpO
2
values.
While on the finger, do not press the 950 against any surface
and do not squeeze or hold the 950. The external spring
provides the correct pressure on the finger; additional
pressure may affect measurements.
Perfusion
The finger must have perfusion adequate to cause the
Perfusion Indicator to blink green in order to give
accurate readings.
Nail Polish
Fingernail polish may reduce light transmission which may
affect SpO
2
values.
In addition, the following factors may degrade performance:
an arterial catheter moisture in the 950
a blood pressure cuff a finger outside the 0.3-1.0” size range
an electrosurgical procedure venous pulsations
an infusion line anemia or a low hemoglobin count
cardiovascular dyes the 950 not at heart level