User Manual for PC-80D Easy ECG Monitor
29
Appendix2 Measurement Result Description
No.
ECG Measurement Result Descriptions
Notes: (bpm
time/minute)
1
No irregularity found.
60 bpm ≤ HR ≤ 100 bpm, No irregular HR found.
2 Suspected a little fast beat
100 bpm < HR <= 110 bpm
3 Suspected fast beat
HR> 110 bpm
4 Suspected missing Beat No heart beat is detected within the period of twice of
average R-R interval
5 Suspected arrest
No heartbeat is detected within 4 seconds
6 Suspected short run of fast beat
More than 3 consecutive premature beats are found
7 Suspected a little slow beat
50 bpm <= HR < 60 bpm
8 Suspected slow beat
HR < 50 bpm
9 Suspected occasional short beat interval
Premature beat
10 Suspected occasional irregular beat
interval
Bigeminy and Trigeminy.
11
Suspected a little fast and interval with
occasional short beat interval
A little fast beat and suspected Premature beat.